ISX-3
Sciospec
The Sciospec EIT16 is Sciospec’s standard EIT system. It’s a very compact yet powerful instrument with 16 channels. With its outstanding performance the EIT16 has become one of the most popular EIT systems for research and industry. The Sciospec EIT32 is a scalable EIT platform that allows for instruments with 32, 64, 128, 256 or even more channels.
The system is designed for measurements for electrical impedance tomography or general multi-port impedance measurement applications. True parallel measurements and tightly synchronized switching allows for fast frame rates.
And still the platform is as flexible and customizable as all our instruments. Configure your instrument to perfectly fit to your application by adding medical grade safety features such as a medical grade power suppy or on isolated housing or just choose the WIFI option for wireless communication with the system.
Type
Electrical Impedance Tomography (EIT) System
Dimensions - WxDxH / cm
24.867 x 18.3 x 9.71
Weight / kg
2.5
Power supply
12VDC (typ.), 18W (max), Barrel connector panel jack (outside diameter 10.5mm, inside diameter 2.1mm) (PN: HEBL 21)
Storage conditions
temperature range: -25 °C … 80 °C, relative humidity: ≤ 80% (noncondensing), The temperature gradient should not exceed 1 K/min to reach operating conditions.
Form Factor
Benchtop / Lab
Channels
≥ 1, ≥ 32, ≥ 64, 128, ≥ 4, ≥ 2
PC interface
USB
Sent directly to Sciospec
Sent directly to Sciospec
Input impedance
≈ 1 TΩ, 2.3 pF
Operation conditions
temperature range: 0 °C ... 40°C, relative humidity: ≤ 80% (noncondensing), altitude: 0 m ... 3000 m
Max. Potential range (±) / V
5
Compliance voltage / ±V
11
Required Compliance Voltage
≥ 10 V, ≥ 5 V, ≥ 2 V
Max. current (output) / A
±50 mA
EIS Capability
Yes
EIS frequency range
100 Hz to 1 MHz
Max. EIS frequency range / MHz
1
AC frequency resolution
40μHz @ f<10kHz, 2mHz @ 10kHz < f <500kHz, 220mHz @ f>500kHz
AC frequency precision
±100 ppm (@ 25 °C) additional max. ±10 ppm over specified temperature range (e.g., at 0 °C) additional ±5 ppm during first year
Galvanostatic EIS Amplitude
10 μA ... 10 mA (peak amplitude)
Galvanostatic EIS Amplitude Resolution
2.81 μA